BRAND | ProMinent |
Product | 792957/OZE 3-mA-2ppm |
Description | . |
Internal code | IMP369001 |
Weight | 0.322 |
Custom code | 90278930 |
Technical specification | Sensor for measuring ozone in clear water. Applications are eg Applications in drinking water and swimming pools Measured variable: ozone (O3) Measurement range: 0.02... 2.0 mg/l Reference method: DPD4 pH range: 4.0... 11.0 Temperature: 5... 40°C max. pressure: 1.0 bar Inflow: 30... 60 l/h (in bypass, open outlet of the measurement water) Output signal: 4... 20 mA ~ measurement range temperature-compensated, uncalibrated, no galvanic separation Sensor assembly: DGM or DLG III Measurement principle: amperometric, 2 electrodes, membrane-covered Installation in the bypass to the process water pipe Connection to the controller via LiYY 2x0.25 sqmm control cable Order number: 725122 (sold by the meter) (Test line not included) Note: **************** For installation in the DLG III flow sensor, a Assembly kit order number 815079 required. (not included) |
default.send_request_amp_Group2