BRAND | Isotech |
Product | DakkS-Vergleichskalibrierung |
Description | DAkkS comparison calibration |
Internal code | ONR4712138 |
Technical specification | DAkkS calibration of a resistance thermometer on 5 temperature points (spread over the calibration range or according to customer requirements). It can be the following Measurement uncertainties are achieved (depending on examinee): -196 °C 15 mK -80 °C to -60 °C 15 mK -60 °C to 90 °C 10 mK 90 °C to 200 °C 30 mK 200 °C to 500 °C 40 mK 500 °C to 660 °C 50 mK 660 °C to 962 °C 150 mK Before calibration, the resistance thermometer is Aged enough. Self-heating and the Hysteresis of the resistance thermometer is performed before Start of calibration examined and when specified which takes into account calibration results. Calibration points: -80°C, 0°C, 230°C, 420°C 660°C |